What is Semiconductor Detector – Definition

A semiconductor detector is a radiation detector which is based on a semiconductor, such as silicon or germanium to measure the effect of incident charged particles or photons. Radiation Dosimetry
silicon strip detector - semiconductors
Silicin Strip Detector Source: micronsemiconductor.co.uk

A semiconductor detector is a radiation detector which is based on a semiconductor, such as silicon or germanium to measure the effect of incident charged particles or photons. In general, semiconductors are materials, inorganic or organic, which have the ability to control their conduction depending on chemical structure, temperature, illumination, and presence of dopants. The name semiconductor comes from the fact that these materials  have an electrical conductivity between that of a metal, like copper, gold, etc. and an insulator, such as glass. They have an energy gap less than 4eV (about 1eV). In solid-state physics, this energy gap or band gap is an energy range between valence band and conduction band where electron states are forbidden. In contrast to conductors, electrons in a semiconductor must obtain energy (e.g. from ionizing radiation) to cross the band gap and to reach the conduction band.

Semiconductor detectors are very similar in operation as photovoltaic panels that generate electric current. In a similar way a current can be induced by ionizing radiation.  As ionizing radiation enters the semiconductor, it interacts with the semiconductor material. It may excite an electron out of its energy level and consequently leave a hole. This process is known as electron–hole pair generation. In semiconductor detectors, the fundamental information carriers are these electron-hole pairs, which are produced along the path taken by the charged particle (primary or secondary) through the detector. By collecting electron-hole pairs, the detection signal is formed and recorded.

Semiconductor detectors are widely used in radiation protection, assay of radioactive materials and physics research because they have some unique features, can be made inexpensively yet with good efficiency, and can measure both the intensity and the energy of incident radiation. These detectors are employed to measure the energy of the radiation and for identification of particles. Of the available semiconductor materials, silicon is mainly used for charged particle detectors (especially for tracking charged particles) and soft X-ray detectors while germanium is widely used for gamma ray spectroscopy. A large, clean and almost perfect semiconductor is ideal as a counter for radioactivity. However, it is difficult to make large crystals with sufficient purity. The semiconductor detectors have, therefore, low efficiency, but they do give a very precise measure of the energy. Semiconductor detectors, especially germanium based detectors, are most commonly used where a very good energy resolution is required. In order to achieve maximum efficiency the detectors must operate at the very low temperatures of liquid nitrogen (-196°C). Therefore, the drawback is that semiconductor detectors are much more expensive than other detectors and require sophisticated cooling to reduce leakage currents (noise).

Types of Semiconductors

Types of Semiconductors

There are many semiconductors in nature and others synthesized in laboratories; however, the best known are silicon (Si) and germanium (Ge). Types of semiconductors:

  • semiconductor detectors - table of parametersSilicon. Silicon is a chemical element with atomic number 14 which means there are 14 protons and 14 electrons in the atomic structure. The chemical symbol for Silicon is Si. Silicon is a hard and brittle crystalline solid with a blue-grey metallic lustre, it is a tetravalent metalloid and semiconductor. Silicon is mainly used for charged particle detectors (especially for tracking charged particles) and soft X-ray detectors. The large band-gap energy (Egap= 1.12 eV) allows us to operate the detector at room temperature, but cooling is prefered to reduce noise. Silicon based detectors are very important in high-energy physics. Since silicon-based detectors are very good for tracking charged particles, they constitute a substantial part of detection system at the LHC in CERN.
  • Germanium. Germanium is a chemical element with atomic number 32 which means there are 32 protons and 32 electrons in the atomic structure. The chemical symbol for Germanium is Ge. Germanium is a lustrous, hard, grayish-white metalloid in the carbon group, chemically similar to its group neighbors tin and silicon. Pure germanium is a semiconductor with an appearance similar to elemental silicon.  Germanium is widely used for gamma ray spectroscopy. In gamma spectroscopy, germanium is preferred due to its atomic number being much higher than silicon and which increases the probability of gamma ray interaction. Germanium is more used than silicon for radiation detection because the average energy necessary to create an electron-hole pair is 3.6 eV for silicon and 2.9 eV for germanium, which provides the latter a better resolution in energy. On the other hand, germanium has a small band gap energy (Egap = 0.67 eV), which requires to operate the detector at cryogenic temperatures.
  • Diamond. Diamond is a solid form of the element carbon with its atoms arranged in a crystal structure called diamond cubic. Diamonds are also very good electrical insulators which strangely is both useful and problematic for electrical devices. Diamond is a wide-band gap semiconductor (Egap= 5.47 eV) with high potential as an electronic device material in many devices. Diamond detectors have many similarities with silicon detectors, but are expected to offer significant advantages, in particular a high radiation hardness and very low drift currents.
  • CdTe and CdZnTe. Cadmium telluride (CdTe) and cadmium zinc telluride (CdZnTe) have been regarded as promising semiconductor materials for hard X-ray and gamma ray detection. The high atomic number and the high density of these materials mean they can effectively attenuate X-rays and gamma rays with energies of greater than 20 keV that traditional silicon-based sensors are unable to detect. This significantly increases their quantum efficiency in comparison with silicon-based . The large band-gap energy (Egap= 1.44 eV) allows us to operate the detector at room temperature. On the other hand, a considerable amount of charge loss in these detectors produces a reduced energy resolution.

Principle of Operation of Semiconductor Detectors  

The operation of semiconductor detectors is  summarized in the following points:

  • Ionizing radiation enters the sensitive volume of the detector and interacts with the semiconductor material.
  • Particle passing through the detector ionizes the atoms of semiconductor, producing the electron-hole pairs. The number of electron-hole pairs is proportional to the energy of the radiation to the semiconductor. As a result, a number of electrons are transferred from the valence band to the conduction band, and an equal number of holes are created in the valence band.
  • Under the influence of an electric field, electrons and holes travel to the electrodes, where they result in a pulse that can be measured in an outer circuit,
  • This pulse carries information about the energy of the original incident radiation. The number of such pulses per unit time also gives information about the intensity of the radiation.

The energy required to produce electron-hole-pairs is very low compared to the energy required to produce paired ions in a gaseous ionization detector. In semiconductor detectors the statistical variation of the pulse height is smaller and the energy resolution is higher. As the electrons travel fast, the time resolution is also very good. Compared with gaseous ionization detectors, the density of a semiconductor detector is very high, and charged particles of high energy can give off their energy in a semiconductor of relatively small dimensions.

Silicon-based Semiconductor Detectors

Silicon-based semiconductor detectors are mainly used for charged particle detectors (especially for tracking charged particles) and soft X-ray detectors while germanium is widely used for gamma ray spectroscopy. A large, clean and almost perfect semiconductor is ideal as a counter for radioactivity. However, it is difficult to make large crystals with sufficient purity. The semiconductor detectors have, therefore, low efficiency, but they do give a very precise measure of the energy. Detectors based on silicon have sufficiently low noise even by room temperature. This is caused by the large band gap of silicon (Egap= 1.12 eV), which allows us to operate the detector at room temperature, but cooling is prefered to reduce noise. The drawback is that silicon detectors are much more expensive than cloud chambers or wire chambers and require sophisticated cooling to reduce leakage currents (noise). They also suffer degradation over time from radiation, however this can be greatly reduced thanks to the Lazarus effect.

Application of Silicon Detectors

Since silicon-based detectors are very good for tracking charged particles, they constitute a substantial part of detection system at the LHC in CERN. Most silicon particle detectors work, in principle, by doping narrow (usually around 100 micrometers wide) strips of silicon to turn them into diodes, which are then reverse biased. As charged particles pass through these strips, they cause small ionization currents that can be detected and measured. Arranging thousands of these detectors around a collision point in a particle accelerator can yield an accurate picture of what paths particles take. For example, the Inner Tracking System (ITS) of a Large Ion Collider Experiment (ALICE) contains three layers of silicon-based detectors:

  • Silicon Pixel Detector (SPD)
  • Silicon Drift Detector (SDD)
  • Silicon Strip Detector (SSD)

Silicon Strip Detectors

Silicon-based detectors are very good for tracking charged particles. A silicon strip detector is an arrangement of strip like shaped implants acting as charge collecting electrodes.

Silicon strip detectors 5 x 5 cm2 in area are quite common and are used in series (just like planes of MWPCs) to determine charged-particle trajectories to position-accuracies of the order of several μm in the transverse direction.  Placed on a low doped fully depleted silicon wafer these implants form a one-dimensional array of diodes. By connecting each of the metalized strips to a charge sensitive amplifier a position sensitive detector is built. Two dimensional position measurements can be achieved by applying an additional strip like doping on the wafer backside by use of a double sided technology.  Such devices can be used to measure small impact parameters and thereby determine whether some charged particle originated from a primary collision or was the decay product of a primary particle that traveled a small distance from the original interaction, and then decayed.

Silicon strip detectors constitute a substantial part of detection system at the LHC in CERN. Most silicon particle detectors work, in principle, by doping narrow (usually around 100 micrometers wide) strips of silicon to turn them into diodes, which are then reverse biased. As charged particles pass through these strips, they cause small ionization currents that can be detected and measured. Arranging thousands of these detectors around a collision point in a particle accelerator can yield an accurate picture of what paths particles take.

For example, the Inner Tracking System (ITS) of a Large Ion Collider Experiment (ALICE) contains three layers of silicon-based detectors:

  • Silicon Pixel Detector (SPD)
  • Silicon Drift Detector (SDD)
  • Silicon Strip Detector (SSD)

Delta E – E Detector – Telescope

delta E - E Telescope - chart
Example of histogram from ΔE-E detector. Each hyperbola-like curve represents different particle in the beam.

In experimental physics, ΔE-E detectors, known as telescopes, are powerful devices for charged particles identification. In order to provide charged-particle identification, telescopes consisting of pairs of thin and thick surface-barrier detectors can be used. These detectors must be positioned in series. The velocity is deduced from the stopping power measured in the thin detectors (ΔE detectors). There is a strong correlation between the energy deposited in each detector. This correlation depends on the mass (A), the charge (Z) and the kinetic energy (E) of each particle. The mass is deduced from the range or from the total kinetic energy loss in the thicker detector (E detector).

Telescopes can be  composed of several detectors (ionization chambers, silicon detectors and scintillators for instance) stacked in order to slow down incident charged particles, the first detector being the thinnest and the last one the thickest. CsI scintillation counters can be, for example used as final E counters. As an example of telescope, an assembly based on two front ΔE silicon detectors (10 or 30 µm) and an E silicon counter 1500 µm thick may be used for detection of high-energy charged particles.

Germanium-based Semiconductor Detectors

Germanium-based semiconductor detectors are most commonly used where a very good energy resolution is required, especially for gamma spectroscopy, as well as x-ray spectroscopy. In gamma spectroscopy, germanium is preferred due to its atomic number being much higher than silicon and which increases the probability of gamma ray interaction. Moreover, germanium has lower average energy necessary to create an electron-hole pair, which is 3.6 eV for silicon and 2.9 eV for germanium. This also provides the latter a better resolution in energy. A large, clean and almost perfect germanium semiconductor is ideal as a counter for radioactivity. However, it is difficult and expensive to make large crystals with sufficient purity. While silicon-based detectors cannot be thicker than a few millimeters, germanium can have a depleted, sensitive thickness of centimeters, and therefore can be used as a total absorption detector for gamma rays up to few MeV.

On the other hand, in order to achieve maximum efficiency the detectors must operate at the very low temperatures of liquid nitrogen (-196°C), because at room temperatures the noise caused by thermal excitation is very high.

Since germanium detectors produce the highest resolution commonly available today, they are used to measure radiation in a variety of applications including personnel and environmental monitoring for radioactive contamination, medical applications, radiometric assay, nuclear security and nuclear plant safety.

Application of Germanium Detectors – Gamma Spectroscopy

As was written, the study and analysis of gamma ray spectra for scientific and technical use is called gamma spectroscopy, and gamma ray spectrometers are the instruments which observe and collect such data. A gamma ray spectrometer (GRS) is a sophisticated device for measuring the energy distribution of gamma radiation. For the measurement of gamma rays above several hundred keV, there are two detector categories of major importance, inorganic scintillators as NaI(Tl) and semiconductor detectors. In the previous articles, we described the gamma spectroscopy using scintillation detector, which consists of a suitable scintillator crystal, a photomultiplier tube, and a circuit for measuring the height of the pulses produced by the photomultiplier. The advantages of a scintillation counter are its efficiency (large size and high density) and the high precision and counting rates that are possible. Due to the high atomic number of iodine, a large number of all interactions will result in complete absorption of gamma-ray energy, so the photo fraction will be high.

HPGe Detector - Germanium
HPGe detector with LN2 cryostat Source: canberra.com

But if a perfect energy resolution is required, we have to use germanium-based detector, such as the HPGe detector. Germanium-based semiconductor detectors are most commonly used where a very good energy resolution is required, especially for gamma spectroscopy, as well as x-ray spectroscopy. In gamma spectroscopy, germanium is preferred due to its atomic number being much higher than silicon and which increases the probability of gamma ray interaction. Moreover, germanium has lower average energy necessary to create an electron-hole pair, which is 3.6 eV for silicon and 2.9 eV for germanium. This also provides the latter a better resolution in energy. The FWHM (full width at half maximum) for germanium detectors is a function of energy. For a 1.3 MeV photon, the FWHM is 2.1 keV, which is very low.

References:

Radiation Protection:

  1. Knoll, Glenn F., Radiation Detection and Measurement 4th Edition, Wiley, 8/2010. ISBN-13: 978-0470131480.
  2. Stabin, Michael G., Radiation Protection and Dosimetry: An Introduction to Health Physics, Springer, 10/2010. ISBN-13: 978-1441923912.
  3. Martin, James E., Physics for Radiation Protection 3rd Edition, Wiley-VCH, 4/2013. ISBN-13: 978-3527411764.
  4. U.S.NRC, NUCLEAR REACTOR CONCEPTS
  5. U.S. Department of Energy, Instrumantation and Control. DOE Fundamentals Handbook, Volume 2 of 2. June 1992.

Nuclear and Reactor Physics:

  1. J. R. Lamarsh, Introduction to Nuclear Reactor Theory, 2nd ed., Addison-Wesley, Reading, MA (1983).
  2. J. R. Lamarsh, A. J. Baratta, Introduction to Nuclear Engineering, 3d ed., Prentice-Hall, 2001, ISBN: 0-201-82498-1.
  3. W. M. Stacey, Nuclear Reactor Physics, John Wiley & Sons, 2001, ISBN: 0- 471-39127-1.
  4. Glasstone, Sesonske. Nuclear Reactor Engineering: Reactor Systems Engineering, Springer; 4th edition, 1994, ISBN: 978-0412985317
  5. W.S.C. Williams. Nuclear and Particle Physics. Clarendon Press; 1 edition, 1991, ISBN: 978-0198520467
  6. G.R.Keepin. Physics of Nuclear Kinetics. Addison-Wesley Pub. Co; 1st edition, 1965
  7. Robert Reed Burn, Introduction to Nuclear Reactor Operation, 1988.
  8. U.S. Department of Energy, Nuclear Physics and Reactor Theory. DOE Fundamentals Handbook, Volume 1 and 2. January 1993.
  9. Paul Reuss, Neutron Physics. EDP Sciences, 2008. ISBN: 978-2759800414.